| First Catalog | 1967 |
|---|---|
| First year price | $3600 |
| Last Catalog | 1973 |
| Last year price | $4295 |
The HP 3460B is a full five digit DVM with an inline nixie tube readout and manual, automatic, or remote range selection.
DC voltage ranges are:
+/-1.00000 V
+/-10.0000 V
+/-100.000 V
+/-1000.00 V; Overranging 20% on all.
Input resistance is 10 MΩ on all ranges; increases to >1 GΩ on 1 and 10 V ranges when displayed value is within 5% of actual measured value (null)
Guard terminal may be as great as +/-500 V from chassis ground; Low input terminal may be as great as +/- 50 V from Guard terminal voltage.
Power: 115 or 230 V +/-10%, 50~60 Hz, 60 W
Dimensions: 16-3/4“ W x 5” H x 19-3/8“ D (425 x 127 x 492 mm); Weight: 38 lbs/17.6 kg
| Available Options | |
|---|---|
| Option 001 | 1224 BCD output |
| Option 002 | 1248 BCD output (Compatible w/ 3461A AC/Ohms Converter), +$150 |
| Option 003 | 1224 BCD output (Compatible w/ 3461A AC/Ohms Converter), +$150 |
| Option 004 | 1248 BCD output, AC Normal-mode rejection filter, +$250 |
| Option 005 | 1224 BCD output, AC Normal-mode rejection filter, +$250 |
| Option 006 | 1248 BCD output, AC Normal-mode rejection filter, (Compatible w/ 3461A AC/Ohms Converter), +$400 |
| Option 007 | 1224 BCD output, AC Normal-mode rejection filter, (Compatible w/ 3461A AC/Ohms Converter), +$400 |